Renishaw Raman Spectrometer
- Make: Renishaw
- Model: InVia
- Serial Number: H41751
- Date of manufacture: April 2006
- Current service engineer contact: Luis Cepeda
- Laser wavelength: 514 nm
Vender supplied Training pdfs
all pdfs zipped together
Standard Operating Procedure (SOP)
last updated 7/29/2021 : SOP
Service History
- 2019 laser replaced
- 7/13/2021 Dichroic mirror assembly replaced and system aligned
Laser Power at amps/watts
Instrument Notes
- use 3 knobs on mirror inside lower left of instrument to get laser in crosshairs in the video image, as of 7/29/21 center of crosshairs are incorrect (about 430 counts when laser at top of bottom center crosshair, static scan default settings)
- Laser's readings on standby: current: 3.3 A power = 0 mW
- Laser's readings on run (100 mW range): current = 7.2 A power = 11.1 mW
- microscope: all levers on upper right hand side are in the maximum up position
- static measurement: grating fixed, smaller spectral width but fast acquistion
- extended measurement: grating rotates to increase spectral width but this slows data acquisition, must have a minimum exposure time of 10 seconds